How to measure blood pressure – Omron Healthcare INTELLISENSE HEM-907XL User Manual

Page 24

background image

24

5. Place the stethoscope on the

patient’s arm.

6. Push the START Button to start the

measurement.

• Do not push the START Button without

wrapping the cuff.

• Do not squeeze or press the cuff during

the measurement.

• If you want to inflate again after the start

of deflation, push the START Button.

• If you want to accelerate deflation after the

start of deflation, push the DEFLATION
(deflation control) / Measurement Results
Display Switch Button. Each time the
Button is pushed, cuff is deflated rapidly in
increments of 5 to 10 mmHg.

9. Push the ON/OFF (power) Button to

turn off the power.

7. Take the readings.

8. Push the STOP Button to remove air

inside the cuff.

• The unit does not automatically deflate in

the MANU Mode.

If the monitor determines that the pressure
value is not correct, an error display appears
(Er1 to 9). In this case, refer to Page 27 and
start the measurement again.

HOW TO MEASURE

BLOOD PRESSURE

(IN MANUAL MODE)