List of error codes – Omron Healthcare INTELLISENSE HEM-907XL User Manual

Page 27

background image

27

LIST OF ERROR CODES

Error code

Explanation

Inflation error

• When the pressure does not exceed

12 mmHg within the set time after the
start of inflation

• When the inflation does not reach the set

cuff pressure within the specified time
after the start of inflation

Deflation error

• When the deflation speed is too fast during

the measurement

• When the deflation speed is too slow

during the measurement

• When the measurement does not finish

within the specified time after starting the
measurement

Overpressure error

• The cuff pressure exceeded 299 mmHg.

Insufficient inflation error

• Blood pressure could not be measured

due to insufficient inflation level.

Indeterminable blood pressure error

• Blood pressure could not be measured

even when the cuff pressure reached the
specified pressure.

Low pulse level error

• Pulse wave was too small.

Blood pressure error

• Relationship between systolic and dias-

tolic pressures was abnormal.

Pulse rate error

• Pulse rate did not stay within the range of

30 to 199 beats/min.

Device error

• Main unit malfunction.

How to correct

• Confirm that the air tube connecting the cuff and

the main unit is connected securely.

• Confirm that the air flow in the air tube

connecting the cuff and the main unit isn’t
being restricted.

• Confirm that the cuff is wrapped correctly (refer

to pages 13 and 14).

• Check bladder for leaks and, if necessary,

replace the bladder with new one (option).

• Confirm that air flow in the air tube connecting

the cuff and the main unit isn’t being restricted.

• If the measurement is made by setting the P-SET

to “AUTO”, ask the patient not to move during
the inflation.

• Confirm that the P-SET is securely set to

“AUTO”. Turn the Knob counterclockwise as far
as it goes until you can hear a click sound.

• If the measurement is made by manual inflation

level setting, set the value to 30 to 40 mmHg
higher.

• Confirm that the cuff is wrapped correctly (refer

to pages 13 and 14).

• Confirm that the cuff is wrapped correctly (refer

to pages 13 and 14).

• Ask the patient not to move during the measure-

ment.

• Check the patient for arrhythmia.

• Contact Omron Healthcare’s Cusomer Service

toll-free at 1-877-216-1336.