When Error Indication Symbols Appear

The following special error symbols will appear in the display if the unit is being used incorrectly:

ERROR SYMBOL

CAUSE

CORRECTION

1 . Cuff underinflated

1) Release air pressure in cuff

2) Wait 10 minutes before taking

another reading .

3) Repeat “Instructions for Use”

steps except inflate cuff 30

mmHg higher than previous

inflation value .

2 . Movement during

measurement

Residual air pressure

in cuff .

1) Release air pressure in cuff

2) Wait 10 minutes before taking

another reading .

3) Repeat “Instructions for Use”

steps

Cuff overinflated (more

than 300 mmHg) .

No pulse or blood

pressure reading due to:

1) Motion

2) Very weak artery

oscillation