beautypg.com

Figure a-2: optical charts, A-2 raynger 3i series operator’s manual – RayTek 3i User Manual

Page 56

background image

A-2

Raynger 3i Series Operator’s Manual

FOCUS POINT D:S = 75:1 @15 m (50 ft) FAR FIELD D:S = 68:1

15

10

20

30

5

40

3

6

9

1.5

12

50

0

0

1.3

2.0

5.0

3.5

8.0

33

51

124

6.5

86

165

203

DISTANCE: SENSOR TO OBJECT IN METERS

DISTANCE: SENSOR TO OBJECT IN FEET

IR Spot Diameter at Lens = 23 mm (0.9 in)
Laser Diameter at Lens = 40 mm (1.6 in)

IR Beam Profile

Laser Profile

LTDL2
LTDL3

Spot Diameter (in)

Spot Diameter (mm)

1ML2
1ML3
1MSC

1.5

0.8

0.5

2.3

3.2

80

57

18

13

4.0

36

100

FOCUS POINT D:S = 180:1 FAR FIELD D:S = 140:1

25 mm @ 4.6 m

1.0 IN @ 15 FT

Focus Point

IR Spot Diameter at Lens = 7 mm (0.3 in)

15

10

20

30

5

40

3

6

9

1.5

12

50

0

0

DISTANCE: SENSOR TO OBJECT IN METERS

DISTANCE: SENSOR TO OBJECT IN FEET

Spot Diameter (in)

Spot Diameter (mm)

LTCL2
LTCL3

5000

3000

1500

1000

36

60

120

180

0

0

0.9

1.9

4.6

7.3

23

48

118

FOCUS POINT D:S = 75:1 FAR FIELD D:S = 20:1

0.3 IN @ 24 IN

8 mm @ 610 mm

Focus Point

IR Spot Diameter at Lens = 23 mm (0.9 in)
Laser Diameter at Lens = 40 mm (1.6 in)

IR Beam Profile

Laser Profile

203

DISTANCE: SENSOR TO OBJECT IN INCHES

DISTANCE: SENSOR TO OBJECT IN MILLIMETERS

Spot Diameter (in)

Spot Diameter (mm)

OPTICAL CHARTS

Note that the optical resolution values are stated at minimum 90% energy (95% for 1M).

G5SC

1.0

1.3

Focus Point

1.5 IN @ 75 IN

2.9

4.9

27

32

38 mm @ 1900 mm

74

137

5000

3000

1500

1000

36

60

120

180

0

0

FOCUS POINT D:S = 50:1 FAR FIELD D:S = 30:1

DISTANCE: SENSOR TO OBJECT IN INCHES

DISTANCE: SENSOR TO OBJECT IN MILLIMETERS

Spot Diameter (in)

Spot Diameter (mm)

IR Spot Diameter at Lens = 23 mm (0.9 in)

Figure A-2: Optical Charts