Raynger 3i Series Operator’s Manual

2-25

Notes:

1. Indicates HAL is active and temperature HAL value.
2. Mode changed to MIN.
3. Mode changed to DIF.
4. Mode changed to AVG.
5. Main and Average temperatures went negative.
6. Average temperature goes positive and is 100°F.
7. Average temperature is up to 999.9°F.
8. Average temperature is now greater than 999.9°F so decimal point is no longer

displayed.

9. Indicates LAL is active and temperature LAL value.
10. Scale is changed from °F to °C.
11. Emissivity is changed to 1.00.

2.6.1.2 Data Output—SET Loop

After entering the SET loop (SET icon activated), setup parameters can be sent out by
pressing the SET button again. This information can be sent out at any time while still
in the SET loop. Unactivated setpoints are printed in lowercase (e.g., hal) and activat-
ed setpoints are printed in uppercase (e.g., HAL). Hi Alarm, Low Alarm, T. Ambient,
Temperature Scale, and DOI are sent out and terminated by a carriage return and line
feed. A total of 39 characters are sent out. The SET values will be printed any time
you press the SET button. Two examples follow:

hal 0085 lal 0065 tam 0100 F;DOI 3600

(hal, lal, tam not activated)

HAL 0085 lal 0065 tam 0100 F;DOI 3600

(HAL activated)

The above lines indicate Hi Alarm = 85, Low Alarm = 65, T. Ambient = 100, the tem-
perature scale is °F, and DOI = 3600 seconds. (The and do not print; these
denote the carriage return and line feed.)

Note: DOI does not apply to the SETUP mode.