Figure 2-11: setup loop–log values – RayTek 3i User Manual
Page 33
Raynger 3i Series Operator’s Manual
2-21
7. Set the TAM value by pressing the ▲ and ▼ buttons.
8. Press the ACTIVATE button to turn on the ambient temperature compensation
function. The Tamb icon will be displayed indicating that it is active. (Pressing
ACTIVATE again turns off the function.)
9. Pull trigger and reset the emissivity to the proper value for the target. You can
now take the target’s temperature using normal procedures. The current tempera-
ture and all computed values, as well as the analog and digital outputs, will now
be based on the compensated measurement.
Note: TAM only affects readings in the RUN loop; LOG loop readings are not affected.
2.4.7.2 SETUP Values for LOG
Figure 2-11 illustrates SETUP for the LOG loop.
To setup values for the LOG modes and functions, do the following:
1. Release or unlock the trigger, if necessary.
2. Press the SET button. The SET icon will be activated.
3. Press the RUN/LOG button, if necessary, so that the LOG icon is activated.
4. Press the ▲ or ▼ buttons to select the LOG location number (LOC mode).
MODE
RUN
LOG
SET
MODE
•
RECALL
ACTIVATE
•
LOCK
LASER
LOG
RCL
SET
°C
°F
HI ( ) LO ( ) Tamb
Press the
▲
and
▼
buttons to change the
Log Location (LOC)
number.
You can select and
log data in any location
number from 1 to 100.
Press the Set button
again, if the display
goes out, to continue
the SETUP functions
RUN
LOG
SET
MODE
•
RECALL
ACTIVATE
•
LOCK
LASER
LOG
RCL
SET
°C
°F
HI ( )
LO ( ) Tamb
Turn High Alarm
on or off with the
Activate button
Change the High Alarm
temperature value with
the
▲
and
▼
buttons
LOC Mode
HAL Mode
TRIGGER
Released & Unlocked
BUTTON
SET & LOG
LOOP ICON
LOG
SET
RUN
LOG
RECALL
•
MODE
ACTIVATE
•
LOCK
LASER
RUN
LOG
RECALL
•
MODE
ACTIVATE
•
LOCK
LASER
Figure 2-11: SETUP Loop–LOG Values