2.6.1

Digital Output

The following sections describe the digital data outputs for each mode.

2.6.1.1 Data Output—RUN Loop

In the RUN loop, the temperature scale, current temperature, emissivity, mode tem-
perature (MAX, MIN, DIF, or AVG), HAL or LAL, a carriage return, and a line feed
are sent out at intervals determined by the DOI (set in the SET loop) when the trigger
is pulled. HAL or LAL is sent out if the temperature is above HAL or below LAL
(when they are activated). Three blank spaces replace HAL or LAL if they do not
meet this condition. There are 5 characters for the mode temperature. The fifth char-
acter is always a space for MAX, MIN, or DIF. For AVG, the fifth character is a space
if the average temperature is 1000° or above. Average temperatures up to 999.9° use
all 5 characters. A total of 29 characters are sent out.

Table 2-2 shows a sample output. The headings at the top of the table explain the con-
tents of each column (these are not part of the output).

2-24

Raynger 3i Series Operator’s Manual

Table 2-2: RUN Mode Output Example

F 0072 E 0 . 95 MAX 0072
F 0073 E 0 . 95 MAX 0073
F 0090 E 0 . 95 MAX 0090
F 0070 E 0 . 95 MAX 0090
F 0071 E 0 . 95 MAX 0090 HAL
F 0081 E 0 . 95 MAX 0090
F 0080 E 0 . 95 MAX 0090
F 0070 E 0 . 95 MIN

0070

F 0070 E 0 . 95 DIF

0020

F 0081 E 0 . 95 AVG

088.4

F –018 E 0 . 95 AVG

–11.5

F 0421 E 0 . 95 AVG

100.0

F 1480 E 0 . 95 AVG

999.9

F 1900 E 0 . 95 AVG 1000
F 0071 E 0 . 95 MAX 0090
F 0081 E 0 . 95 MIN

–018 LAL

F 0081 E 0 . 95 MIN

–018

C

0027 E 0 . 95 MIN

–028

C

0027 E 1 . 00 MIN –028

or °F Temp Emissivity Value Mode Temperature 3 Blank Spaces Return

Feed

Main

E for

Emissivity

Mode

HAL, LAL, or

Carriage

Line

S

p

a

c

e

S

p

a

c

e

S

p

a

c

e

S

p

a

c

e

S

p

a

c

e

S

p

a

c

e

°C

Notes

1

2
3
4
5
6
7
8

9

10
11

(see below)