13

VPAP Adapt SV System

HumidAire™

You will need a medium size 21" (52 cm) air tube to connect the

VPAP Adapt SV

unit to the

HumidAire™ (see Figure 7).

1

Make sure both the

HumidAire and the VPAP Adapt SV are turned off.

Fill the

HumidAire with water as described in the humidifier manual. Place the

filled water chamber inside the

HumidAire.

2

Gently detach enough of the tube clips to release the end of the pressure sensor
tube from the

VPAP Adapt SV air tubing.

Connect the

VPAP Adapt SV air tubing to the left connector port on the humidifier,

and the medium 21" (52 cm) air tubing to the right connector port. Close the
HumidAire lid.

3

Place the

VPAP Adapt SV on top of the HumidAire. Never place the

VPAP Adapt SV unit underneath the humidifier. (This is to avoid water spilling into
the unit.)

4

Connect the free end of the medium air tubing to the

VPAP Adapt SV.

Connect the pressure sensor tube by twisting the Luer lock onto the
VPAP Adapt SV (see page 10).

HumidAire

Medium air tubing

VPAP Adapt SV air tubing

Pressure sensor tube