ResMed VPAP Adapt SV System User Manual
Page 16

10
2
Connect air tubing 
Connect the air tube firmly onto the air outlet at the front of the unit.
Figure 4: Front view of VPAP Adapt SV
To attach the pressure sensor tube to the unit, screw the Luer connector onto 
the socket on the side of the unit.
Figure 5: Twisting the Luer lock connectors together
!
WARNING
Use care when handling the air tubing. Be particularly careful when 
attaching the Luer connector to the VPAP Adapt SV. If you kink or 
excessively twist the sensor tube it may affect the protection offered by the 
mask pressure alarm system.
3
Attach the mask to the air tubing.
The mask system should already be assembled. If not, please refer to the user 
instructions supplied with the mask. 
Connect the mask system to the free end of the air tubing.
Figure 6: Assembling the VPAP Adapt SV system
Air tubing connector
Pressure sensor tube 
connector
Proximal cuff