ResMed VPAP Adapt SV System User Manual
Page 16
10
2
Connect air tubing
Connect the air tube firmly onto the air outlet at the front of the unit.
Figure 4: Front view of VPAP Adapt SV
To attach the pressure sensor tube to the unit, screw the Luer connector onto
the socket on the side of the unit.
Figure 5: Twisting the Luer lock connectors together
!
WARNING
Use care when handling the air tubing. Be particularly careful when
attaching the Luer connector to the VPAP Adapt SV. If you kink or
excessively twist the sensor tube it may affect the protection offered by the
mask pressure alarm system.
3
Attach the mask to the air tubing.
The mask system should already be assembled. If not, please refer to the user
instructions supplied with the mask.
Connect the mask system to the free end of the air tubing.
Figure 6: Assembling the VPAP Adapt SV system
Air tubing connector
Pressure sensor tube
connector
Proximal cuff