Humidifiers

A humidifier may be required if you are experiencing dryness of the nose, throat,
or mouth. The following ResMed humidifiers are compatible with the
VPAP Adapt SV:

Figure 7: Humidifiers compatible with the VPAP Adapt SV

When using a humidifier, stop the

VPAP Adapt SV by using the Start/Stop key or

by removing the mask to trigger the SmartStop function. Do not switch the
VPAP Adapt SV off at the power switch or the power source while it is running, as
this could result in condensation in the motor.

Please refer to the relevant humidifier user manual for detailed instructions and
warnings.

!

CAUTIONS

Be very careful not to allow water to enter the VPAP Adapt SV as this
could affect therapy and/or damage the device.

Do not tip the humidifier while it is connected to the VPAP Adapt SV. If
water enters the VPAP Adapt SV device, turn off the main power switch
at the back of the unit and unplug it from the power outlet. Return the
unit to your equipment supplier for checking.

Connecting a Humidifier

Note: After connecting or removing a humidifier, always perform the Air Circuit
Learn function (see page 22).

!

WARNING

When using a humidifier, position it so that it is lower than your head will
be during sleep, and at the same level or lower than the VPAP Adapt SV.

HumidAire 2i™

heated humidifier

Medium air tubing

21" (52 cm)

Humidifier accessory (HumidAire only)

HumidAire™

heated humidifier