beautypg.com

Output characteristics, Pxi_star trigger characteristics – National Instruments NI PXIe-6672 User Manual

Page 39

background image

Appendix A

Specifications

NI PXIe-6672 User Manual

A-4

ni.com

Output Characteristics

Frequency range .....................................DC to 105 MHz

Output impedance...................................50

Ω, nominal

Output coupling ......................................DC

Voltage level...........................................0 to 1.6 V into 50

Ω;

0 to 3.3 V into open circuit,
typical

Absolute maximum applied voltage

1

......±5.25 V, max

PXI_CLK10 synchronized trigger clock
to out time, t

CtoQ

2

....................................10.7 ns, typical

Output-to-output skew, synchronous......500 ps, typical

PXI_STAR Trigger Characteristics

PXI_STAR <0..16> to
PXI_STAR <0..16> output skew
at NI PXIe-6672 backplane connector....300 ps

3

, typical

Asynchronous delays, t

pd

PXI_STAR <0..16> to
PFI <0..5> output.............................13 to 17 ns, typical

PXI_STAR <0..16> to
PXI_TRIG <0..7> output.................18 to 24 ns, typical

1

Stresses beyond those listed can cause permanent damage to the device. Exposure to absolute maximum rated conditions for

extended periods of time can affect device reliability. Functional operation of the device outside the conditions indicated in
the operational parts of the specifications is not implied.

2

Relative to PXI_CLK10 at backplane connector.

3

This specification applies to all synchronous routes to the PXI_Star lines, as well as asynchronous routes from the PFI inputs

to the PXI_Star lines.