2. Before Operating the Scanner

2.1

Power Supply Precautions

In order to use the ScanTouch 210 safely and correctly, and to prevent
problems, pay careful attention to the following points:

Be sure to use an AC100V-120V/AC 200V - 240V, 50/60Hz power
supply. The power cord should be changed according to the power
voltage.

When using a power voltage of more than AC 125V

Use a suitable power cord compliant with the safety standards of the
country in which it is used, with a plug of AC 250V, 15A rating
(NEMA 6P-15), and insulation of at least SVT type and more than
AWG18 in thickness.

When using a power voltage of AC 125V or less

Use a suitable power cord compliant with the safety standards of the
country in which it is used, with a plug of AC 125V, 10A rating and
insulation of at least SVT type and more than AWG18 in thickness.

Be sure that the electrical outlet of the power supply is grounded.
Carry out the grounding in common with the other machines it is
being connected with. Unless common grounding is used, a ground
loop will occur, which will cause electric shock and noise static.

* The shape of the plug depends on the country of use.