Differential Connections for Ground-Referenced
Signal Sources

Figure 2-6 shows how to connect a ground-referenced signal source to a
channel on the NI 783xR/784xR/785xR configured in DIFF input mode.

Figure 2-6. Differential Input Connections for Ground-Referenced Signals

With this connection type, the instrumentation amplifier rejects both the
common-mode noise in the signal and the ground potential difference
between the signal source and the NI 783xR/784xR/785xR ground, shown
as V

cm

in Figure 2-6. In addition, the instrumentation amplifier can reject

common-mode noise pickup in the leads connecting the signal sources to
the device. The instrumentation amplifier can reject common-mode signals
when V+

in

and V–

in

(input signals) are both within their specified input

ranges. Refer to the NI R Series Intelligent DAQ Specifications, available at

ni.com/manuals

, for more information about input ranges.

DIFF Input Mode Selected

+

+

+

+

V

cm

V

s

I/O Connector

AISENSE

AIGND

V

m

AI+

AI–

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

Measured

Voltage

Instrumentation

Amplifier