Channel Trigger
Trigger modes and appendant registers
Channel window trigger for entering signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower level define a win-
dow. Every time the signal enters the the window from the
outside, a triggerevent will be detected.
Channel window trigger for leaving signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower level define a win-
dow. Every time the signal leaves the the window from the
inside, a triggerevent will be detected.
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXWINENTER
10040
SPC_HIGHLEVEL0
42000
r/w
Sets the window’s upper level relatively to the channel’s input range.
board dependant
SPC_LOWLEVEL0
42100
r/w
Sets the window’s lower level relatively to the channel’s input range.
board dependant
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXWINLEAVE
10050
SPC_HIGHLEVEL0
42000
r/w
Sets the window’s upper level relatively to the channel’s input range.
board dependant
SPC_LOWLEVEL0
42100
r/w
Sets the window’s lower level relatively to the channel’s input range.
board dependant