Trigger modes and appendant registers
Channel Trigger
(c) Spectrum GmbH
67
Channel pulsewidth trigger for short positive pulses
The analog input is continuously sampled with the selected
sample rate. If the programmed triggerlevel is crossed by
the channel’s signal from lower to higher values (rising
edge) the pulsewidth counter is started. If the pulsewidth
counter reaches the programmed amount of samples, no
trigger will be detected.
If the signal does cross the triggerlevel again within the the
programmed pulsewidth time, a triggerevent will be detec-
ted.
Channel pulsewidth trigger for short negative pulses
The analog input is continuously sampled with the selected
sample rate. If the programmed triggerlevel is crossed by
the channel’s signal from higher to lower values (falling
edge) the pulsewidth counter is started. If the pulsewidth
counter reaches the programmed amount of samples, no
trigger will be detected.
If the signal does cross the triggerlevel again within the the
programmed pulsewidth time, a triggerevent will be detec-
ted.
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXPOSG_SP
10002
SPC_HIGHLEVEL0
42000
r/w
Set it to the desired triggerlevel relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXNEG_SP
10012
SPC_HIGHLEVEL0
42000
r/w
Set it to the desired triggerlevel relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255