Guidance and manufacturer’s declaration - electromagnetic immunity

The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
UP-DR80MD should assure that it is used in such as environment.

Immunity test

IEC 60601 test

level

Compliance level

Electromagnetic environment-guidance

Electrostatic
discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.

Electrical fast
transient/burst

IEC 61000-4-4

±2 kV for power
supply lines

±1 kV for input/
output lines

±2 kV for power
supply lines

±1 kV for input/
output lines

Mains power quality should be that of a typical commercial or
hospital environment.

Surge

IEC 61000-4-5

±1 kV differential
mode

±2 kV common
mode

±1 kV differential
mode

±2 kV common
mode

Mains power quality should be that of a typical commercial or
hospital environment.

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines

IEC 61000-4-11

< 5% U

T

(> 95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(> 95% dip in U

T

)

for 5 sec

< 5% U

T

(> 95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(> 95% dip in U

T

)

for 5 sec

Mains power quality should be that of a typical commercial or
hospital environment. If the user of the UP-DR80MD requires
continued operation during power mains interruptions, it is
recommended that the UP-DR80MD be powered from an
uninterruptible power supply or a battery.

Power frequency
(50/60 Hz)
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at least
characteristic of a typical location in a typical commercial or
hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.