OPERATING INSTRUCTIONS

Control panel

Temperature display

The unit is set at the factory to obtain the proper temperature automatically (4

o

C

1

o

C).

The high temperature alarm is preset at 6

o

C. The low temperature alarm is preset at 2

o

C.

Table 1. Temperature display

: Lamp ON

: Lamp OFF

Operation

Key operated

Display

Display mode

1 Plug-in, Power ON

Average monitor bottle

temperature

UPPER

LOWER

Average temperature

2 Press key.


Upper monitor bottle

temperature

UPPER

LOWER

Upper temperature

3 Press key.


Lower monitor bottle

temperature

UPPER

LOWER

Lower temperature

4 Press key.


Average monitor bottle

temperature

UPPER

LOWER

Average temperature

5 Repeat from 2

Note:

The displayed temperature represents the temperature sensed by a thermal sensor attached to the

monitor bottles mounted to the upper and lower portion of the chamber.

The displayed temperature may not always be at 4

o

C. The chamber probes are representative of high

and low average temperatures within the chamber.

Defrost cycles

There is no need for routine defrosting of the unit as this occurs automatically as follows:

1. Cycle defrost

To keep the chamber temperature stable, the refrigeration compressor is cycled on and off. During

“off”

period any frost which has accumulated on the evaporator is melted by energizing a defrost heater. This

will not have any discernible effect on the chamber temperature.

2. Forced defrost

When the ambient humidity is high, or a large amount of damp product is being stored inside the chamber,

there is a possibility that cycle defrost may not be enough to remove all of the frost on the evaporator. In

this case, a forced defrost cycle can be initiated.

When the unit is operating under a forced defrost cycle, the current chamber temperature and

“dF” is

displayed alternately on the digital temperature display.

Once the forced defrost cycle is completed, normal operation resumes.

The chamber temperature of blood bag (400mL) rises about 1 to 2

o

C during the defrosting.