START-UP OF UNIT

Follow the procedures for the initial and consequent operations of the unit.

1. Connect the power cord to the dedicated outlet with appropriate rating.

2. On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the

buzzer stop key (BUZZER). The alarm will operate until the temperature monitor bottle sensor goes into

the range of 4

o

C

2

o

C. See Table 2 in page 17.

3. Ensure that both monitor bottles have been filled with 10% glycerol as outlined earlier in this manual.

4. The unit has been factory preset to operate at 4

o

C

1

o

C. There is no need to set the temperature.

5. Turn on the fluorescent light switch to check the light. After checking, turn off the switch if the light is

not necessary.

6. When you are satisfied that the unit is working correctly, begin slowly placing items into the chamber to

minimize the temperature rise.

Stock of containers

Always distribute items so as not to disturb the air circulation in the chamber. Disruption of the

air flow can cause items to freeze or reduce the uniformity of the chamber temperature.

Never put any articles on the top of the unit.

Intake air vent

Do not block this vent with containers. If this vent

becomes blocked, temperature regulation

becomes unstable. Do not place paper or vinyl

near the vent as they may be sucked into the fan.

Exhaust air vent

Do not block this vent with containers as this can

impede air circulation. Items stored near the

exhaust air vent can freeze.

Load line

In this refrigerator, storage space is limited to allow

the passage of cool air around the chamber. The

label

“load line” is fixed to both sides of the

chamber. Always stock items within this line.

CAUTION

Never store corrosive substances in this unit. This may lead to damage to the inner components or

electric parts.

Load line

Stock space

Fan(inside

Exhaust air vent

Intake air vent

Exhaust air vent

Load line