Troubleshooting – Invacare 3600XL User Manual
Page 33
SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
33
FIGURE 4.1 Calibrating the Disinfectant Flow Meter
Troubleshooting
NOTE: If problems are not remedied by the suggested means above, contact an ICCI dealer or
technical support.
SYMPTOMS
FAULTS
SOLUTION
Hydromassage System
System will not start.
No electrical power to tub.
Check and reset breaker.
System not prepared.
Be sure water level is ABOVE jets.
Check that inlet plug on tub is depressed.
Switch hose disconnected
Reconnect hose to switch and motor.
System is ineffective.
Water deposits in whirlpool jets. Disassemble and clean whirlpool jets.
Whirlpool jets closed.
Open whirlpool jets.
Aerator closed.
Open aerator.
Whirlpool intake clogged.
Inspect intake and remove obstruction.
Tub Operation
No water flow.
Water supply turned off.
Turn on water supply.
Restricted water flow.
RADA valve. Refer to RADA
owner’s manual
Inspect and clean RADA valve.
Disinfectant Cabinet Door
Floating
Ball
Knob
Flow
Meter
Disinfectant Valve
DETAIL “B”
Screw
Injector
NOTE: Model 3650 shown. Models 3600 and 3600XL have the same contents in the disinfecting cabinet.
DETAIL “A”
Burp Fitting
Burp Fitting
Models 3600 and 3600XL
Model 3650
Disinfectant Siphon Tube
DETAIL
“C”