Number of
flashes
Fault
Impact on
Scooter
Notes
4
Current Limit
Time Out
Drive inhibited
•
The scooter has drawn too much current
for too long, possibly because the motor
has been over-worked, jammed or
stalled. Turn the scooter power off, leave
it off for a few minutes, and then turn the
power back on again.
•
The controller has detected a short-
circuited motor. Check the cable harness
for short and check the motor: Contact
your authorised Invacare® Dealer.
5
Brake Fault
Drive inhibited
•
Check that the declutching lever is in the
engaged position.
•
The park brake coil or wiring is faulty.
Check the park brake and wiring for open
or short circuits. Contact your authorised
Invacare® Dealer.
6
Out Of Neutral
At Power Up
Drive inhibited
•
Throttle is not in neutral position when
turning key switch on. Return throttle to
neutral, turn power off, and back on
again.
•
Throttle may need to be re-calibrated.
Contact your authorised Invacare®
Dealer.
7
Speed Pot
Error
Drive inhibited
•
The throttle or its wiring may be faulty or
incorrectly set up. Contact your
authorised Invacare® Dealer.
8
Motor Volts
Error
Drive inhibited
•
The motor or its wiring is faulty. Contact
your authorised Invacare® Dealer.