SECTION 5—MAINTENANCE

Part No 1143215

27

BAR750

SECTION 5—MAINTENANCE

Maintenance Checklist

Invacare recommends the following maintenance procedures be conducted at least once 
every three years. This form is provided as a guide to help with documentation.

Mechanical Inspection and Maintenance

❑ Inspect all bed components for damage or excessive wear.
❑ Visually examine all welds for cracks.
❑ Inspect the head and foot spring sections for bending, warping or damage.
❑ Check drive shaft and drive shaft connections for bending, damage or excessive wear.
❑ Inspect pull tubes and mounting hardware for bending, damage or excessive wear.
❑ Inspect all bolts and rivets to ensure that they are securely tightened and functioning 

properly.

❑ Check casters to ensure they lock, if applicable, and roll properly.
❑ Ensure casters are free of debris.

Electrical Inspection and Maintenance

❑ Inspect all electrical bed components for damage or excessive wear (i.e., cracked or 

broken housings, or worn components).

❑ Check pendant, power and motor cords for chafing, cuts or excessive wear.
❑ Make sure all plugs are fully attached and free of damage.
❑ Check all functions (head, foot, and bed ends raise and lower properly).

Pass/Fail

❑ PASS:Bed and all components have passed all of the above inspection criteria and the 

bed is ready for placement into a client’s home.

❑ FAIL:Bed, or any component, has failed any of the above inspection criteria. Tag the 

bed or component with a complete description of the failure(s) and have the bed 
serviced.