beautypg.com

IDEAL INDUSTRIES 61-320 User Manual

Page 14

background image

Page 12

Table 1 AC Voltage Test:

Step

Input

Frequency

Reading

1 5.800V

50Hz

5.743

to

5.857

2 5.800V

500Hz

5.743

to

5.857

3 58.00V

50Hz

57.43

to

58.57

4 58.00V

500Hz

57.43

to

58.57

5 580.0V

50Hz

574.3

to

585.7

6 580.0V

500Hz

574.3

to

585.7

7 750V

50Hz

738

to

762

8 750V

500Hz

738

to

762

6. Turn the rotary switch to “V ” position.
7. Set the calibrator for the voltage from step 1 to 6 in Table 2.
8. Compare the reading on the meter display with the display reading shown in Table 2.
9. If the display reading falls outside of the range shown in Table 2, the meter does not meet specification.

Table 2 DC Voltage Test:

Step

Input

Reading

1 580.0mV

576.9

to

583.1

2

-580.0mV

–576.9 to –583.1

3 5.800V

5.769

to

5.831

4 58.00V

57.69

to

58.31

5 580.0V

576.9

to

583.1

6 900V

894

to

906


Testing the Resistance Function
To verify the accuracy of the resistance function, do the following:
1. Connect the calibrator to VΩHz and COM on the meter.
2. Turn the rotary switch to Ω.
3. Apply the inputs for step 1-6 in Table 3.
4. Compare the meter display readings to the display readings in Table 3.
5. If the display reading falls outside of the range shown in Table 3, the meter does not meet specification.

Table 3 Ω Resistance Test:

Step

Source

Reading

1 580.0

Ω 575.7

to

584.3

2 5.800K

Ω 5.757

to

5.843

3 58.00K

Ω 57.57

to

58.43

4 580.0K

Ω 575.7

to

584.3

5 5.800M

Ω 5.740

to

5.860

6 58.00M

Ω 57.11

to

58.59

* Lead resistance on the 400Ω range is not included in error.


Form number TM61320-2-4

Rev 2 September 2002

This manual is related to the following products: