Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with
equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.

Table B. Required Equipment

Equipment

Required Characteristics

Recommended Model

Calibrator


AC Voltage Range:
0-750 VAC
Accuracy: ±0.07% (Basic)
Frequency Range: 40 ~ 1KHz
Accuracy: ±2%

DC Voltage Range: 0-1000V DC
Accuracy: ±0.006% (Basic)

Current Range: 0 ~ 10A
Accuracy: AC (40Hz to 1KHz): ±0.08% (Basic)
DC: ±0.02% (Basic)

Frequency Source: 5.00Hz ~ 100MHz
Accuracy: ±0.001%

Amplitude: 0.5V p-p ~ 1.0V p-p (square wave)
Accuracy: ±5%

Resistance range: 1Ω ~ 100M Ω
Accuracy: ±0.03% (Basic)

Capacitance Range: 1PF ~ 10mF
Accuracy: ±0.10% (Basic)


Fluke 5500 or Wavetek
9100 Calibrator or
equivalent


Form number TM61320-2-4

Rev 2 September 2002