beautypg.com

IDEAL INDUSTRIES 61-320 User Manual

Page 12

background image

Page 10

Testing Fuses
To test the internal fuses of the meter:
1. Turn the rotary selector switch to the Ω position.
2. To test FS1, plug a test lead into VΩHz input terminal and touch the probe to the mA input terminal. The
display should indicate between 0.0 to 0.2 Ω. FS1 (1A 600V - Bussmann BBS-1 recommended). If display
reads higher than 0.2Ω, replace the fuse.

Fuse Replacement
Refer to Figure 1B to replace the fuse:

Figure 1B


Use only a fuse with the amperage, interrupt, voltage, and speed rating specified.
Fuse rating: 1A, 600V, Fast

Replacing Fuses

Warning

To avoid electrical shock, remove the test leads and any input signals before replacing the
battery or fuses. To prevent damage or injury, INSTALL ONLY quick acting fuses with the
following Amp/Volt current interrupt rating:
FS1 Fuse: 1A, 600V, FAST. Minimum interrupt rating: 10,000A.

Cleaning

Warning

To avoid electrical shock or damage to the meter, never allow water inside the case. To avoid
damaging the meter’s housing, never apply solvents to the meter.


Form number TM61320-2-4

Rev 2 September 2002

This manual is related to the following products: