Maintence and Troubleshooting
Unipro 3.5
Rev. 11.00
December 30, 1995
The SET display messages are shown below:
Message
Description
Z-A
Zero input A
Z-B
Zero input B
Z-C
Zero input C
Z-SW
Zero slide wire (from
input C)
S-A
Span input A
S-B
Span input B
S-C
Span input C
S-SW
Span slide wire (from
input C)
NOTE
It is very important to be sure the SET display
is showing the proper mode before making an
adjustment or the wrong value will be changed.
Adjustment Sensitivity
The adjustment sensitivity works in the
following manner. If the digit farthest right in
the process display is flashing, then each press
of the [Up Arrow] or [Down Arrow] key will
change the calibration value shown by one
unit. This is the least sensitive position. If the
digit farthest left is flashing, then each press of