For Serial Program Mode, all commands must be framed by an STX (02 Hex) and ETX (03 Hex).
To recall defaults:
1) Transmit 999999 through the Serial Port. This will put the scanner in serial

program mode. Scanning will be suspended and the scanner will respond with an ACK (06 Hex).

2) Transmit 999998 through the Serial Port. This is the Recall Defaults bar code in the

MetroSelect guide. The scanner will respond with an ACK (06 Hex).

3) Transmit 999999 through the Serial Port. This will cause the scanner to exit

program mode and save the new settings. The scanner will beep 3 times and send an
ACK (06 Hex).

If at anytime, the scanner cannot recognize a command, it will respond with a NAK (15 Hex).
Please refer to the Installation and User’s Guide of the scanner for a complete description of Serial
Program Mode.