Metrologic Instruments StratosS MS2220 User Manual
Page 50
44
S
CALE
O
PERATION
C
ALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display
7. Go
to Group 2 Step 6 (Calibration) of the Service Mode. Press the Tare ( ) key to advance to Group 2, then use the
Zero ( ) key to advance to step 6 of group 2.
8. Press the Tare ( ) key to change the prompt to ‘YES’.
9. Press the Zero ( ) key to accept.
10. The display will now show “- - - - - -“. This indicates the scale is ready to take a zero reading. Check the platter to ensure
that it is seated properly and nothing is touching or interfering with its freedom to move.
11. Press the Zero ( ) key to accept and begin taking the Zero Capacity calibration reading.
If the scale is unable to capture a stable reading it will not store any value and advance to the next step.
Verify nothing is touching or interfering with the scale and/or platter.
12. Once the calibration reading has been successfully taken, the display will now show “FULCAP” (Full Capacity). Place the
Full Capacity weight, either 15kg or 30lb (whichever matches the scales selected Weight Unit), on the center of the scale
platter. Wait at least 3 seconds for the scale to stabilize.
13. Press the Zero ( ) key to accept and begin taking the Full Capacity calibration reading.
If the scale is unable to capture a stable reading it will not store any value and advance to the next step.
Verify nothing is touching or interfering with the scale and/or platter.
14. Once the calibration reading has been successfully taken, the display will now show “UNLOAD” (Zero Capacity).
Remove all weight from the platter at this time. Wait at least 3 seconds for the scale to stabilize.
15. Press the Zero ( ) key to accept and begin taking the final Zero Capacity calibration reading.
16. The display will now begin the initialization sequence and return to the normal weighing mode in the weight units
calibrated.
17. The unit’s calibration must now be verified as required by state and/or local Weight and Measures regulations
(starting on Page 49).
Need to Start Over?
If for any reason you need to restart the calibration process, restart at Step 1 on page 42.
* These procedures are for MS2220 StratosS Scanner/Diva Scale models only
.