F-801 Internal RO System by GE Osmonics

12552e – 27 Feb 04

4- 32

Routine Maintenance

4.6.2 BIOSAN Storage Procedure

WARNING:
It is recommended that gloves (latex/vinyl) and eye protection be worn when
handling BIOSAN.

1.

Follow the BIOSAN procedure in this chapter.

2.

Do not rinse the RO out after BIOSAN has been circulated in the RO.

3.

Disconnect the disinfect line, waste line, feed water line, power and product
line.

4.

Coil the hoses and power cord, and cover the RO with a large plastic bag to
protect the RO from dust.

WARNING:
After storage, the RO unit must be completely rinsed of BIOSAN, and a bacterial
culture performed and interpreted before use. Disinfect RO unit if necessary.
Perform AAMI analysis if it has been one year or more since last done.
CAUTION:
The F-801 RO may be stored for up to two weeks in BIOSAN. After two weeks,
if longer storage is needed, the RO may be rinsed and re-packed with BIOSAN.