F-801 Internal RO System by GE Osmonics

12552e – 27 Feb 04

1 - 12

General Information

1.10 Specifications:
1.10.1 Input Water Requirements

Minimum

Maximum

Input

flow

Rate 2.0

GPM

N/A

Temperature

(water)

10°C

(50°F)

32°C

(90°F)

NOTE:
Operating at lower temperatures may make it impossible for the F-801 RO to
produce adequate permeate for the dialysis machine.

pH (operating)

5.5

10.5

TF

Membranes

pH w/chloramines present (operating) 5.5

8.5

TF

Membranes

pH

(cleaning)

2.0

11.0

TF

Membranes

Pressure,

dynamic

30

PSI

100

PSI

1.10.2 Flow

Rates

Product Flow based on the following conditions:
New

Membranes

Operating

Pressure:

150

PSI

Product

Pressure:

1

PSI

Feed Water Temperature:

77

°

F (25

°

C)


GPM LPM GPD LPD
.41 1.55 588 2225.8

Membrane flux may vary ±15%
Waste Flow: 2000ml/min (

±

10%)


1.10.3 Electrical

Requirements

Motor

1/3 HP

Voltage 115

V~

Hertz

60

Amps

10

Phase

single

Power Cord (length)

8 ft (2.4 m)