Symptom
Carbonator pump not turning on.
Carbonator pump is not turning off.
Carbonation system is not
carbonating.
Carbonation system is producing
excessively carbonated beverages.
Possible cause
1. Loose connection.
2. Electrode connections on
carbonator tank are satisfied.
3. Carbonator pump or level
control board is defective.
1. Loose connection.
1. Loose connection.
2. CO
2
pressure.
3. Water pressure is too high.
4. Carbonator pump is defective.
1. CO
2
pressure is set too high.
2. Drinks may be warm.
Solution
1. Check carbonator pump power at
plug and at level control board.
2a. Check connection between low
level electrode and level control
board.
2b. Check ground wire on carbonator
tank.
3. Replace pump/control board.
1a. Check connection between high
level electrode and level control
board.
1b. Replace electrodes/control board.
1. Check all CO
2
and water
connections.
2. Set CO
2
pressure to 75 psi.
3. Lower water pressure to
approximately 50 psi.
4. Replace pump.
1. Set CO
2
pressure to 75 psi.
2. See “warm drinks” symptom.
If problems persist after following this basic troubleshooting guide, call Follett’s technical
service department at (800) 523-9361 or (610) 252-7301.