Unpacking and installation – Iris Sample Processing StatSpin® Express 2 Primary Tube Centrifuge User Manual
Page 5

1
Section
1
Unpacking and Installation
Inspect Packaging
The Express 2 and its accessories are delivered in one carton. If the centrifuge or accessories have
suffered any damage in transport, please inform your carrier immediately.
NOTE: Save shipping carton and components to simplify return should service be required.
Verify Contents
The package contains:
One Express 2 centrifuge (Model No. M501-22)
One Centrifuge Rotor (Product No. RTX4)
Four Rotor Inserts for 13x75 mm tubes (Product No. SV01)
One Universal Switching Power Supply
(StatSpin Product No. 01-3553-001, APS Product No.
AD-740U-1240)
One grounded line cord (for North American use only)
One Operator’s Manual
One Tool for Removing Rotor, 7/16” Nut Driver
Install System
1. Place the Express 2 on a level surface suitable for laboratory instrumentation.
2. Maintain a 300mm clearance boundary around the centrifuge for ventilation and safety.
3. Position the Express 2 away from direct sunlight and sources of heat or cold.
(See Appendix A for specifications.)
4. Remove blue foam packaging around the rotor before operating the centrifuge.
Connect Power
Plug the power supply into a grounded outlet supplying the voltage and frequency indicated on the
power supply. When power is connected, the 30 s. LED will illuminate and the cover lock will
release.
WARNING - North American Installation:
Only use the power supply included with the unit.
Use of other power supplies or transformers will damage the Express 2 electronics and void
the warranty.
WARNING - Outside of North America:
Do Not Use the Line Cord Supplied. Use power
cord for at least 1.0 Amp or more with an IEC320/CEE22 female connector and male
connector suitable for the power outlet to be used.