How to use this manual – Iris Sample Processing StatSpin® Express 2 Primary Tube Centrifuge User Manual
Page 4

ii
How to use this manual
This manual along with information contained on product labels and in package inserts should
provide you with all the information you need to operate and maintain the Express 2.
Notes appear in italics to highlight information. When the information requires special attention, a
caution symbol appears next to the italicized text.
Please pay close attention to the instructions that accompany the notes and symbols as well as the
standard laboratory practices outlined by your facility and local regulatory agencies. The table below
lists all the CAUTIONS/WARNINGS for the Express 2.
WARNING - North American Installation: Only use the power supply included
with the unit. Use of other power supplies or transformers will damage the Express
2 electronics and void the warranty.
WARNING - Outside of North America: Do Not Use the Line Cord Supplied. Use
power cord for at least 1.0 Amp or more with an IEC320/CEE22 female connector
and male connector suitable for the power outlet to be used.
CAUTION - The cover interlock bypass is designed for emergency use only. If the
equipment is not used properly, safety may be impaired.
CAUTION- The rotor (RTX4) has a finite lifespan that is dependant on usage.
Cracks or swelling that develop indicate the rotor should be replaced immediately.
CAUTION – Failure to properly install the rotor may result in damage to the
centrifuge and will void the warranty.
CAUTION – Never operate the Express 2 without the rotor in place.
CAUTION - Replace balance tube every 10 cycles to prevent possible breakage
due to the repeated stress of centrifugation.
CAUTION
–
Unplug the Express 2 from the wall outlet before performing
maintenance.
WARNING – Do not expose Express 2 and its rotor to strong or concentrated acids,
bases, esters, aromatic or halogenated hydrocarbons, ketones or strong oxidizing
agents.
CAUTION – DO NOT spray the bowl or outer surfaces with detergent or bleach.
Excess liquid will harm the electronics and subsequent problems may not be
covered under warranty.
CAUTION - Universal Precautions should be followed on all specimens, regardless
of whether a specimen is known to contain an infectious agent. (See references)
Please use the system as intended. Improper use of the Express 2 Primary Tube Centrifuge and/or
its accessories may cause damage to the system, inaccurate results, or potentially nullify warranties.
The Express 2 Primary Tube Centrifuge, and its associated components are covered under US
patent nos. 4,846,974 & 4,981,585 & 5,566,919