Iris Sample Processing StatSpin® Express 2 Primary Tube Centrifuge User Manual
Page 12

8
Troubleshooting
No Power Light
↓
Is the power cord completely
plugged into the wall?
→ NO→
Plug power cord into the wall.
↓
YES
↓
Is the wall outlet working properly?
→ NO→
Try another wall outlet.
↓
Is the barrel connector from the
transformer completely plugged into
the centrifuge?
→ NO→
Plug the barrel connector into the centrifuge.
↓
YES
↓
Contact StatSpin or an Authorized
Repair Facility.
Centrifuge Will Not Spin/Shuts
Off Prematurely
↓
Is the cover completely closed and
cannot be opened manually?
→ NO→
Completely close the cover.
↓
YES
↓
Contact StatSpin or an Authorized
Repair Facility.
Centrifuge will not open at end
of cycle
↓
Is the bowl gasket completely
seated and sealing the bowl?
→ NO→
Press the bowl gasket all the way down into
the housing slot.
↓
YES
↓
Contact StatSpin or an Authorized
Repair Facility.