beautypg.com

Hudson Robotics Micro10x User Manual

Page 11

background image

11

Figure 2.3.1


Volume – Any integer value from 1 to 9999uL may be selected.

Speed – This adjusts the percentage of the pumps maximum speed. More viscous
liquids will require a slower pump speed to avoid cavitation.

Depth – Negative values are the distance below the travel height where the tips
will move during the dispensing, allowing the user to dispense with the tips in the
wells. Positive values will move the dispensing location above the travel height.
Entering 0 will keep the travel height and the dispense height the same.

Height – This is the height the tips will travel above the nest during the
dispensing. This value should be at least 1mm above the measured plate height.
The Standard Microtiter plate is ~15mm.

Wells – Select from 96, 384 or 1536