System specifications, Hardware specifications, Software specifications – Dako Autostainer User Manual
Page 51: Hardware specifications software specifications
Section 12
System Specifications
Hardware Specifications
Dako Autostainer
Dimensions
1.02m W x 0.69m D x 0.61m H
40" W x 27" D x 24" H
Weight
140 lb. (63.5 kg)
Electrical Requirements
120V 110/120V (±10%) 60Hz (±2Hz)
220V 220/240V (±10%) 50Hz (±2Hz)
Normal Operating Temperature 18°C – 26°C (64°F – 79°F)
Slide Rack Capacity
1 – 12 glass slides
Total Slide Capacity
1 – 48 glass slides
Reagent capacity
64 different reagents (15 mL/reagent vial)
Reagent dispense volumes
100, 150, 200, 400 and 600 µL
Reagent probe volume capacity 100 µL minimum, 1200 µL maximum
Computer Controller
Dako reserves the right to change the computer controller specifications at any time.
Computer specifications may vary.
Monitor 15”
color
Printer
Color InkJet printer
Surge Protector 120 VAC, 15 Amps, 50/60 Hz
Software Specifications
Operating
System
Windows
®
98, XP or equivalent.
Operating Logic Designed to calculate the most time-efficient sequence to complete a
programmed staining run.
Protocol Logic
Flexible selection of steps including rinses. Maximum 35 protocol
steps (including rinse steps).
Idle Rinse
Buffer is applied to slides during a staining run to keep specimens
wet when no reaction is occurring.
Rinse Default
Settings
Buffer rinse every 30 minutes prior to the start of a run, when delayed
start is selected.
Buffer rinse every 30 minutes during a staining run.
Water rinse every 60 minutes after completion of run.
Default settings for time and the amount of buffer dispensed can be
changed by a Dako representative upon request.
Incubation Time Recommended minimum time is two minutes.
Autostainer Handbook
43