Bio-Rad BioFrac™ Fraction Collector User Manual
Page 95

15-Pin D Connector
User Guide
| 87
Table 23.
15-Pin D connector output signals
Pin #
Output Signals
Type
Active
Wire Color
Description
4
READY
Level
Low
Green
Indicates that the fraction
collector diverter valve is
connected and functioning
properly.
9
10
START REMOTE
DEVICE
Relay
Closed
Green/Black
Orange/Black
Relay used for starting a
remote device such as a
pump. Relay is closed
when Run is pressed and
opened when Stop is
pressed. Active in Local
mode.
11
EVENT MARK
Pulse
Low
Blue/Black
Sends a negative pulse
(200 ms) for each fraction
advance. If the delay
function is being used, the
drophead advance will be
separated from the event
mark by the delay time,
volume, or drops.
15
SIGNAL
GROUND
Blue/White
Signal ground.
All input and output pins, except the relay terminals, are TTL compatible. All input pins are with
10 k
pull-up. The relay pins are used to control low-voltage DC contact-closure type devices,
such as 12 V, 10 mA.
Caution: Do not apply 110 VAC or 220 VAC directly to the relay terminal pins.
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- iMark Microplate Absorbance Reader (112 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- ChromLab™ Software (178 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)