beautypg.com

Technical data, Technical data for remote ignition unit aod 1.2, Technical data for decomposition vessel aod 1.1 – IKA AOD 1 User Manual

Page 66: Technical data for guard aod 1.3, Technical data for oxygen filling station c 48, Technical data -1, Technical data for remote ignition unit aod 1.2 -1, Technical data for decomposition vessel aod 1.1 -1, Technical data for guard aod 1.3 -1, Technical data for oxygen filling station c 48 -1

background image

IKA

-WERKE AOD 1

Ver. 01 01.03

Page 9-1

9 Technical

Data

9.1

Technical Data for Remote Ignition Unit AOD 1.2

Rated voltage/frequency

115 V 50/60 Hz
230 V 50/60 Hz

Power consumption

45 W

Fuse

0.3 A (automatic)

Degree of protection to DIN 40 050

IP 21

Protection class

1 (with protective earth)

Over-voltage category

2

Contamination level

II

Ambient temperature

5 °C ... 40 °C

Max. ambient relative humidity

80 %

Dimensions

135 x 185 x 115 mm (BxDxH)

Weight

2.7 kg

Housing

Sheet metal, painted

9.2

Technical Data for Decomposition Vessel AOD 1.1

Please see the Operating Instructions AOD 1.1 for technical data of the decomposi-
tion vessel.

9.3 Technical

Data

for Guard AOD 1.3

Dimensions

300 x 530 x 320 mm (BxDxH)

Wall thickness

10 mm

Weight

12.2 kg

Length of connection lead

5 m (only connection to Remote Ignition
Unit AOD 1.2)

Application

Suitable for all Type AOD 1.1 decomposition
vessels

9.4

Technical Data for Oxygen Filling Station C 48

Please see the Operating Instructions for the Oxygen Filling Station C 48.