beautypg.com

KYORITSU 6310 Instruction Manual User Manual

Page 296

background image

16.9 Specification of Clamp sensor KEW6310

16.13

KEW6310

MODEL8141

MODEL8142

MODEL8143


Rated current

AC1000mA

Output voltage

AC0 ~ 100mV

(AC100mV/ AC1000mA)

Measuring range

AC0 ~ 1000mA

Accuracy
(sine input)

±1.0%rdg±0.1mV (50/ 60Hz)

±2.0%rdg±0.1mV (40Hz ~ 1kHz)

Phase
characteristics

-------------------

Temp. & humidity
range (guaranteed
accuracy)

23±5°C, relative humidity 85% or less (no condensation)

Operating temp.
range

0 ~ 50°C, relative humidity 85% or less (no condensation)

Storage temp. range

-20 ~ 60°C, relative humidity 85% or less (no condensation)

Allowable input

AC100Arms (50/60Hz)

AC200Arms (50/60Hz)

AC500Arms (50/60Hz)

Output impedance

approx 180Ω approx

200Ω approx

120Ω

Location for use

indoor use, altitude 2000m or less

Applicable standard

IEC 61010-1, IEC 61010-2-032

Measurement CAT.III (300V) Pollution degree 2

IEC61326 (EMC standard)

Withstand voltage

AC3540V / 5 sec

between Jaws - enclosure

between enclosure – output terminal

between Jaws – output terminal

Insulation
resistance

50MΩ or more/ 1000V

between Jaws - enclosure

between enclosure – output terminal

between Jaws – output terminal

Max conductor size

Φ24mm

Φ40mm

Φ68mm

Dimension

100(L) × 60(W) × 26(D)mm

(protrusions are not included)

128(L) × 81(W) × 36(D)mm

(protrusions are not included)

186(L) × 129(W) × 53(D)mm

(protrusions are not included)

Cable length

approx 2m

Output terminal

MINI DIN 6PIN

Weight

approx 150g

approx 240g

approx 490g

Accessory

Instruction manual,

Carrying case

Option 7146

(Φ4 Banana plug),

7185 (Extension lead)