6.1.1 Safety
Comply with:

IEC/EN 61010-1

Insulation: double

insulation

Pollution degree:

2

For inside use, max height:

2000m (6562 ft)

Installation category:

CAT IV 600V, CAT III 1000V to ground


6.1.2 General

data

Mechanical characteristics
Dimensions (L x W x H):

215 x 74 x 43mm ; 8 x 3 x 2 in

Weight (including battery):

285g (10 ounces)

Max conductor size:

30mm (1in)


Supply
Battery type:

1x9V alkaline battery NEDA 1604 IEC 6F22

Low battery indication:

” is displayed when the battery level is too low

AutoPowerOFF

after 15min of idleness (not disabled)


Display
Characteristics:

3½ LCD (max 2000 points) plus decimal point, unit
symbol indication and backlight

Sample rate:

2 times/sec

Conversion mode:

Mean value


6.2 ENVIRONMENTAL

CONDITIONS

6.2.1 Climatic

conditions

Reference temperature:

23° ± 5°C (73°F ± 41°F)

Operating temperature:

5 ÷ 40°C (41°F ÷ 104°F)

Operating humidity:

<80%RH

Storage temperature:

-20 ÷ 60 °C (-4°F ÷ 140°F)

Storage humidity:

<80%RH


This product conforms to the prescriptions of the European directive on low voltage

2006/95/EEC and to EMC directive 2004/108/EEC

6.3 ACCESSORIES
6.3.1 Standard

accessories

The content of a standard package is the following:
• Instrument

HT9012

• Test leads - Cod. 4413-2

• Carrying

bag

• Battery

• User

manual