6 technical specifications – HT instruments HT4010 User Manual
Page 36
HT4010
EN - 15
6 TECHNICAL
SPECIFICATIONS
6.1 CHARACTERISTICS
Accuracy is indicated as [% rdg + dgt]. It is referred to: 23°C ± 5°C with RH <80%RH
DC Voltage
Range Resolution Accuracy
Input
impedance
Overload
protection
200.0mV 0.1mV
±(0.8%rdg + 2dgt)
10M
Ω
600VDC/ACrms
2.000V 0.001V
±(1.5%rdg + 2dgt)
20.00V 0.01V
200.0V 0.1V
600V 1V
±(2.0%rdg + 2dgt)
AC Voltage
Range Resolution
Accuracy
(50
÷ 60Hz)
Input
impedance
Overload
protection
200.0mV 0.1mV
±(1.5%rdg + 3.5mV)
10M
Ω
600VDC/ACrms
2.000V 0.001V
±(1.8%rdg + 8dgt)
20.00V 0.01V
200.0V 0.1V
600V 1V
±(2.5%rdg + 8dgt)
AC Current
Range Resolution Accuracy
Frequency
range
Overload
protection
2.000A 0.001A
±(2.5%rdg + 10dgt)
50
÷60Hz
600Arms
20.00A 0.01A
±(2.5%rdg + 4dgt)
200.0A 0.1A
600A 1A
±(4.0%rdg + 8dgt)
Resistance and Continuity test
Range Resolution Accuracy
Buzzer
Overload
protection
200.0
Ω 0.1Ω
±(1.0%rdg + 4dgt)
≤150
Ω
600VDC/ACrms
2.000k
Ω
0.001k
Ω
±(1.5%rdg + 2dgt)
20.00k
Ω
0.01k
Ω
200.0k
Ω
0.1k
Ω
2.000M
Ω
0.001M
Ω
±(2.5%rdg + 3dgt)
20.00M
Ω
0.01M
Ω
±(3.5%rdg + 5dgt)
Diode test
Features
Test current
Open voltage
0.3mA typical
1.5VDC