Valco electronic Valve Oven User Manual
Page 39

Chapter 3
Operating Valves Automatically
Operation
39
2. Scroll to
Add Run Time Event
and press ENTER.
3. Select
Valve
and press ENTER.
4. Scroll to the sampling valve that you want to add and press ENTER.
The following screen appears:
The
Inject at
parameter is the time into the run that the sample is to be
injected. The
Inject for
parameter is the time that the valve remains into the
inject position. After that time, the valve returns to the Load position. If the
Inject for
time exceeds the GC run time, the GC will automatically reset the
valve to the Load position at the end of the run.
5. Repeat as necessary for any remaining sampling valves. If you need to edit a
run table entry, select the entry to be edited and press ENTER.
NOTE
To delete a run table entry, select it and press CLEAR. The TRACE GC Ultra will ask you if
you really want to delete the entry. Press YES.
SELECT EVENT to add
Signal
Valve
<
External Event
SELECT PARAM to Add
Valve #1
Sampling
<
Valve #2
Switching
Valve #3
Sampling
RUN TIME EVENT
Valve #1
Sampling
Inject at
0.00
Inject for
0.00