beautypg.com

Symptom: probable cause: solution – Invacare IRC10LX User Manual

Page 27

background image

SECTION 7—TROUBLESHOOTING GUIDE

Part No 1118353

27

Platinum

Series

YELLOW or RED
LIGHT Illuminated.

Alarm:
CONTINUOUS
On red light only.

* Only applies to
5LXO

2

and

10LXO

2

SensO

2

models.

1. Low oxygen purity.*

1. Clean or Replace filters.

2. Kinked or blocked tub-
ing, cannula or humidifier.*

2. Inspect for kinks or block-
ages. Correct, clean or replace
item. Once corrected, turn
power OFF for 60 seconds and
then turn power back ON.

3. Flowmeter set too low:
0.5 L/min (or less) for
Platinum 5; 1.0 L/min (or
less) for Platinum 10.

3. Check flowmeter is set to
1.0 L/min. or more.

4. Unit overheating due to
blocked air intake.

4.a. Remove and clean cabinet
filters.
4b. Move concentrator at least
three inches from walls, drap-
eries, and furniture.

5. Insufficient power at
outlet.

5. DO NOT use extension
cords. Move to another elec-
trical outlet or circuit.

6. Internal repairs
required.

6. Call Service Provider.

GREEN LIGHT
with YELLOW
Light Flashing.

1. Internal repairs
required.

1. Call Service Provider.

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

This manual is related to the following products: