Symptom: probable cause: solution – Invacare IRC10LX User Manual
Page 27
SECTION 7—TROUBLESHOOTING GUIDE
27
YELLOW or RED
LIGHT Illuminated.
Alarm:
CONTINUOUS
On red light only.
* Only applies to
5LXO
2
and
10LXO
2
SensO
2
models.
1. Low oxygen purity.*
1. Clean or Replace filters.
2. Kinked or blocked tub-
ing, cannula or humidifier.*
2. Inspect for kinks or block-
ages. Correct, clean or replace
item. Once corrected, turn
power OFF for 60 seconds and
then turn power back ON.
3. Flowmeter set too low:
0.5 L/min (or less) for
Platinum 5; 1.0 L/min (or
less) for Platinum 10.
3. Check flowmeter is set to
1.0 L/min. or more.
4. Unit overheating due to
blocked air intake.
4.a. Remove and clean cabinet
filters.
4b. Move concentrator at least
three inches from walls, drap-
eries, and furniture.
5. Insufficient power at
outlet.
5. DO NOT use extension
cords. Move to another elec-
trical outlet or circuit.
6. Internal repairs
required.
6. Call Service Provider.
GREEN LIGHT
with YELLOW
Light Flashing.
1. Internal repairs
required.
1. Call Service Provider.
SYMPTOM:
PROBABLE CAUSE:
SOLUTION: