Figure 6.2 preventive maintenance record, Series – Invacare IRC10LX User Manual
Page 25

SECTION 6—MAINTENANCE
25
FIGURE 6.2 Preventive Maintenance Record
ON EACH INSPECTION
MODEL NO.
I RC
SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
INVACARE PLATINUM OXYGEN CONCENTRATOR
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
EVERY
4
,380
HOURS FOR IRC5LX (EVERY
2,160
H
OURS FOR IRC10LX)
Check Oxygen Concentration
(UNITS WITHOUT
SensO
2
)
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
(
Check Power Loss Alarm
AS REQUIRED
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End of Compressor
*NOTE:
Refer to Preventive Maintenance section of Service Manual.
NOTE:
2,160 hours are equivalent to usage 24 hours per day, for 90 days.
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Check Oxygen Concentration
UNITS WITH SensO
2
)
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
E
V
ERY 2
6
,280 HOUR
S
OR 3 YEARS, WHICHEVER COMES FIRST