SECTION 4—TYPICAL PRODUCT PARAMETERS
13
SECTION 4—TYPICAL PRODUCT
PARAMETERS
*NOTE: Stated concentration levels achieved after initial warm‐up
period (approximately 30 minutes).
Electrical Requirements:
120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Rated Current Input:
Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Sound Level:
Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Altitude:
Platinum 5 and XL - Up to 6,000 ft (1828
meters) above sea level without degradation
of concentration levels. From 6,000 ft (1828
meters) to 13,129 ft (4000 meters) below
90% efficiency
Platinum 10 - Up to 4,000 ft (1230 meters)
above sea level without degradation of con-
centration levels. From 4,000 ft (1230
meters) to 13,129 ft (4000 meters) below
90% efficiency
*Oxygen Output
Concentration Levels:
All 5LXO2/5LX models 87% to 95.6% for
flows ranging 0.5 to 5 L/min and all
10LXO2/10LX models 87% to 95.6% for
flows ranging 2 L to 10 L/min
Maximum Outlet Pressure:
Platinum 5 and XL: 5 psi ± 0.5 psi (34.5 kPa
± 3.45 kPa)
Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45
kPa)