beautypg.com

Select data file, Print report, Indices – HP CD Leycom CFL-512 User Manual

Page 4: Additional features in conduct nt, Cfl-512 conduct nt – analyse recorded data, Pressure-volume analysis, Volume calibration

background image

CFL-512

Conduct NT – Analyse Recorded Data





Select data file

Retrieve previously acquired data files for subsequent

analysis.

The comment and some file characteristics are displayed in

the list.


Pressure-volume analysis

The end-systolic and end-diastolic PV points are indicated with markers (blue

and green lines). Appropriate (user-selected) functions can be fitted through

these points taken and varying pressures and volumes to calculate contractility

and compliance.


Print Report

Create your own print reports with indices. Select the reports you want to print.
Use the short-cut button

to print the latest report.

Indices

Create your own indices-display by selecting the indices that are available in the

list. Conduct NT has more than 50 parameters in his list of indices.


Volume calibration

Vc calculation (parallel conductance)

Vc will be calculated automatically from the data acquired during a hypertonic

saline injection.

Alpha calculation

To correct the underestimation of true volume, the results measured by the
conductance catheter should be compared to the results of an independent

cardiac output measurement, thermodilution, echo or angiography for example.

Additional features in Conduct NT…..

• Interval measurement

• Several filters can be used

• Exporting data and indices to comma-separated file

• Beat to beat analysis

• Complete file management

• Signals can be individually scaled.

CD Leycom info 0310 V0440