6 hold function setting, Excell precision co., ltd – Excell Precision 9903 Dolphin II User Manual
Page 19

EXCELL PRECISION CO., LTD
.
Dolphin 9903 V 1.1 SME300000041
17
Use the ON/ZERO or UNITS
key to input a parameter
1-6 Hold Function Setting
ON/ZERO
⇒
Upward digit selecting (0-9)
UNITS
⇒
Downward digit selecting (9-0)
TARE
⇒
Move the cursor rightward
NET/GROSS
⇒
Move the cursor leftward
Press the TARE key
Hold function setting
Default setting = 0
0
⇒
No hold function
1
⇒
“peak” hold function mode
2
⇒
“stable” hold function mode 1
3
⇒
“stable” hold function mode 2
The display shows the current hold
function mode
Press the TARE key
Press the TARE key to
complete the function setting
Input “00”
⇒
No Hold function
⇒
“Peak” hold function mode
When the weight keeps changing, the scale will automatically hold
and display the largest value.
Press any key to clear the held value.
⇒
“Stable” hold function mode 1
When the scale becomes stable, the display shows and holds the
weight value, which can not be changed.
Press any key to clear the held value.
⇒
“Stable” hold function 2
When the scale becomes stable, the display shows and holds the
weight value, which can not be changed.
When the scale is re-zeroed (weight less than 10d), the held value is
automatically cleared.