Technical Data
Display:
3 digit LCD, 1999 digit
Duration of measurement:
3-5s
Measurement principle:
current-voltage measurement principle in accordance with
EN61557-5
Measurement voltage:
25V/ 50V selection
Current/power consumption: approx. 10mA/ 0,1W max. 500mA briefly, during measurement
Measurement current:
max. 12mA
Frequency (f1/ f2):
127Hz/ 140Hz, selectable
Measurement method:
2 pole/ 3 pole/ 4 pole
Range
Tolerance
RE:
0.05 - 2.00 Ohm ±(4% + 0.1 Ohm)
±(6% + 0.1 Ohm)*
RE:
2.00 - 1999 Ohm ±(4% + 2 Digit)
±(6% + 3 Digit)*
RH, RS:
0.1 - 50 kOhm
±(10% + 3 Digit)*
IM:
0.1 - 13mA
±(10% + 3 Digit)*
*operating error
Nominal use range:
Correct polarity:
for 3/4 pole measurements
Noise voltage:
0...5V, max. 10% of measurement voltage
aux. earth (RH), probe resis. (RS):
max. 50 kOhm
Ratio RH/ RE:
max. 1000/1
Ratio RS/ RE:
max. 10000/1
Earth resistance (RE):
max. 2 kOhm
Reference range:
Correct polarity:
for 3/4 pole measurements
Noise voltage:
0 V
aux. earth (RH),
probe resis. (RS)
0 Ohm
Earth resistance (RE):
max. 2 kOhm
Auto power off
after approx. 5 min
Power supply
6 x 1.5 V Type IEC LR6 (Mignon/Size AA)
Low Bat at
6.6 V ±5 % (OFF at 6.1V ±5 %)
Battery life
approx. 1 year for normal use
Dimensions (LxWxD)
235x105x68 mm
Weight
approx. 640g (inc. 6 x batteries)