6 Technical specifications
6.1 Technical
features
Accuracy is shown in terms of [% of the reading + number of digits]. It refers to the
following ambient conditions: temperature 73°F +/- 2° (23 °C +/- 1°) with relative humidity
< 75%.
6.1.1 Measuring the insulation.
Accuracy
Range
100 .. <500 V
500 .. <1000 V
1000 .. 2500 V
2500 .. 5000 V
100 K - 500 K
±(20% + 10dgt) ±(20% + 10dgt) ±(20% + 10dgt) ±(20% + 10dgt)
500 K - 50 G
±(5% + 1dgt)
±(5% + 1dgt)
±(5% + 1dgt)
±(5% + 1dgt)
50 G - 100 G
±(20% + 10dgt)
±(5% + 1dgt)
±(5% + 1dgt)
±(5% + 1dgt)
100 G - 250 G
- - -
±(20% + 10dgt)
±(5% + 1dgt)
±(5% + 1dgt)
250 G - 500 G
- - -
- - -
±(20% + 10dgt)
±(5% + 1dgt)
Minimum resistance resolution:
1%
Voltage indication resolution:
1.22 V
6.1.2 Current which can be delivered
Voltage setting
Max current
100 - 500 V
3.00 mA
1000 - 5000 V
1.40 mA
The output current is automatically limited.
6.1.3 Voltage measurement
Voltage range
Resolution
Accuracy
0 - 500 V
1,22V
±(10%+10 dgt)
500 - 5000 V
1,22V
±(10%+ 5 dgt)
The limits shown here are valid both for DC of any polarity and for AC.
Discharge time at 5000V: <1.5s x
µF.
6.1.4 Safety
Instrument according to standard:
EN 61010-1
Insulation:
Class
2,
Double
insulation
Pollution level:
2
Indoors use; max altitude:
6,562 feet (2000 m)
Over-voltage category:
CAT III 600V
Serial port insulation:
25 mm. (max. 5000 V)