Amprobe ACD-30P Clamp-On-Power-Meters User Manual
Page 13

12
�
EMC: en 61326-1 This product complies with requirements of the following
european Community Directives: 89/336/eeC (electromagnetic Compatibility) and
73/23/eeC (low Voltage) as amended by 93/68/eeC (Ce Marking). however, electrical
noise or intense electromagnetic fields in the vicinity of the equipment may disturb the
measurement circuit. Measuring instruments will also respond to unwanted signals that
may be present within the measurement circuit. Users should exercise care and take
appropriate precautions to avoid misleading results when making measurements in the
presence of electronic interference.
Electrical (23 °C ± 5 °C) < 75% Rh
AC Voltage
Voltage
Range
Accuracy
600.0V
50hz to 60hz
± (0.5% rdg + 5d)
45 to 50hz, 60 to 500hz
± (1.5% rdg + 5d)
500hz to 3.1khz
± (2.5% rdg + 5d)
CMRR :
> 60 dB @ DC to 60 hz, Rs = 1 kΩ
input impedance:
2 MΩ, 30 pf nominal
Crest factor:
< 2.3 : 1 at full scale; < 4.6 : 1 at half scale
ACV AutoVA
TM
Threshold:
30VAC (40 to 500 hz) nominal
DC Voltage
Range:
600.0 V
Accuracy:
± (0.5% rdg + 5d)
nMRR :
> 50 dB @ 50/60 hz
CMRR :
>120 dB @ DC, 50/60 hz, Rs = 1 kΩ
input impedance:
2 MΩ, 30 pf nominal
DCV AutoVA
TM
Threshold:
2.4VDC nominal
PEAk-rms HoLD (ACA & ACV only)
Response: 65ms to 90% rdg
ohms
Range:
000.0 to 999.9 Ω
Accuracy:
± (1.0% rdg + 6d)
open Circuit Voltage :
0.4VDC typical
Audible Continuity tester
Audible threshold: between 10Ω and 300Ω.
Response time: 250μs
ACA Current (Clamp-on)