beautypg.com

Using notices, Remote control overview – RIGOL DSA875 User Manual

Page 11

background image

-10-

Using Notices

1. To use the tracking generator output terminal:

CAUTION

To avoid damage to the tracking

generator, the reverse power cannot

exceed +10 dBm when the frequency

less than 10 MHz; the reverse power

cannot exceed +20 dBm when the

frequency more than 10 MHz. The

reverse DC voltage cannot exceed 50 V.

2. To use the RF input terminal:

CAUTION

To avoid damage to the instrument,

for the signal input from the RF input

terminal, the DC voltage component

and the maximum continuous power

of the AC (RF) signal component

cannot exceed 50 V and +20 dBm

respectively.

Remote Control Overview

DSA800 supports communication with PC via USB, LAN or GPIB

(option) interface for remote control. The remote control is

realized on the basis of SCPI (Standard Commands for

Programmable Instruments) command set through two ways:

user-defined programming and PC software (such as

RIGOL

Ultra Sigma). When the instrument is in remote mode, the

icon is displayed in the user interface and the front

panel keys (except Esc) are locked. At this point, you can press

Esc to exit remote mode.

This manual is related to the following products: